3. E-O error message
If this error message remains on screen after turning meter on, the meter has got da-
maged. Please remove the batteries and keep the OK button pressed about 15-20
seconds. Replace the batteries and turn on the meter. If E-0 appears again, please
contact your local distributor.
4. E-2 error message
Strip has been removed from the instrument during measurement.
5. E-3 error message
Used strip is inserted. Repeat measurement with new test strip!
Check-strip is inserted instead of test strip. Replace it with a test strip.
Strip is not correctly inserted.
Blood was applied before strip insertion. Repeat measurement with new test strip!
6. E-5 error message
Strong electromagnetic field (e.g.: mobile phone) or high intensity electrostatic
discharge interfere with the meter. Repeat the measurement.